HOME ResearchLaboratory Equipment High Temperature/Pressure Water Chemistry Loop System
The system is used to simulate the critical operation environments of high-temperature and high-pressure in nuclear and thermal power plants. The levels of dissolved oxygen and dissolved hydrogen are adjusted to simulate the optimal experiment conditions similar to that of a power plant.
System for safe simulation of the condition's of a nuclear (350°C, 15MPa) or thermal (600°C, 25MPa) power plant
Program for realtime measurement and storage of temperature, pressure, and water chemistry conditions (DO, DH, and pH)
Safety system to prevent excessive pressure, overheating, gas leakage, and cooling issues
System with a stable operation for over one year
System for precise and safe control (mass flow controller [MFG]) of argon, hydrogen, and other gases
Class | Description | Remarks |
---|---|---|
Temperature |
Max. 400℃ | Option: 600℃ |
Pressure |
Max. 25MPa | N/A |
DO |
0 ~ 20ppb, 0 ~ 1,000ppm | N/A |
DH |
0 ~ 80cc/kg | Adjustable |
Electrical conductivity |
0.02 ~ 2,000mS/m | Resolution: ±1% |
pH |
1.0 ~ 11.0pH | Resolution : 0.1 |
Customized fabrication available according to the requirements.